1. Ge Oec 9900 Elite

Re: re: OEC 9900 Service Manual December 12, 2012 05:30 Would you please send me the 9900 service manual scottfishman2007@gmail.com let me know what you need and if I have it ill send it to you. GE OEC 9900 Elite Mobile C-Arm - The GE OEC 9900 Elite is designed to deliver superb clinical images during both simple and challenging imaging cases, with fast and easy positioning. See which configuration – based on II size, type of C and optional features - will best meet your clinical needs. Imaging profiles optimize your image to compensate for bone and tissue.

Description

Details

RECONDITIONED OEC 9900 ELITE C-ARM

The reconditioned OEC 9900 Elite select configurations are ideal for Office,Outpatient and Ambulatory Center settings.

Elite

Superb Imaging
• 1k x 1k x 16 bit resolution imaging
• Advanced image processing accentuates what needs to be seen and attenuates what isn’t needed
• Dynamic Range Management technology processes images such as bone and vasculature separately for optimum definition
• Dose management options for more control: Digital Spot,High Level Fluoroscopy, Standard Fluoroscopy, Low Dose, and Pulse Mode
• Smart Metal for enhanced imaging when metal is present
• AutoTrak sampling window automatically moves with the anatomy to ensure superb images during live fluoro

Workflow Efficiency
• Simple user interface with touchscreen reduces procedure time to improve efficiency
• Flat panel monitors on an articulating arm allow viewing of images from anywhere in the room
• Superb mechanical C-arm design features counterbalancing and friction controls for easier AP and lateral movement, orbital rotation, and head-to-toe angulation

Refurbishing Process:
All OEC-9800’s go through a multistep refurbishing process. Each unit is restored to a like-new status. Pain Management, Orthopedics and Rehabilitation Services are just a few of the applications utilizing the refurbished OEC-9800 c-arm. The rotating x-ray tube allows for greater penetration and resolution for the lumbar area when doing needle localizations or other work involving the thicker anatomical areas of patients. For barium swallows a VCR or Digital Video Recorder can be utilized for better record keeping and post study analysis. The 1K X 1K CCD Technology used in the OEC-9800 image system allows for crisp, clear imaging. Whatever the purpose, a refurbished OEC-9800 will deliver the results you can depend on.

9900

Mechanical repairs/cosmetic maintenance
• All defective parts and covers are repaired or replaced
• C-arm is completely disassembled and repainted
• “C” is removed from main frame assembly
• Horizontal arm removed, inspected and repaired as necessary
• Vertical column removed and bearings removed inspected and replaced as required
• X-ray tube removed and inspected for artifacts and internal integrity
• Image intensifier removed and inspected for any imperfections and grid integrity
• Camera removed
• Wheels removed, cleaned and or replaced from main chassis
• Chassis inspected for structural defects and repaired as necessary
• Vertical lift assembly is tested for drift and adjusted/replaced if necessary
• All bearings are inspected, cleaned and re-lubricated, or replaced as needed
• Control panel display is tested and replaced if necessary
• All wheels are inspected and lubricated and/or replaced as needed, with new cable pushers installed
• All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
• C-arm is completely disassembled and repainted
• Steering is tested and adjusted is required
• Foot switch and hand switch provided on each unit
• A complete set of operator and service manuals are provided

Manual

Re-assembly
• All components sent to C-arm reassembly area for rebuilding
• “C”, Image tube and camera, Vertical column, main control and power cable installed
• New batteries are installed
• New CRT monitors installed in monitor cart
• Monitor cart rebuilt
• When C-arm is completely rebuilt mechanically system is sent for calibration and final testing

Testing/calibration
• Image intensifier and X-ray tube are tested for stability and balance
• Image intensifier is tested for resolution and gain to be within OEM specifications
• X-ray tube bearings are tested for noise and coast time
• X-ray tube filaments and stator windings are tested
• X-ray tube radiation output is verified to be within OEM specifications
• X-ray generator high voltage tested
• Maximum dose rate is set in compliance with government regulations
• CCD camera calibrated to OEM specifications
• X-ray beam is aligned for each field size
• C-arm is tested for current leakage requirements
• All monitors are aligned, resized, centered and focused
• All monitors are adjusted for brightness and linearity
• Video system is optimized for gray scale and resolution
• All power supplies are tested and calibrated or replaced as necessary

Oec 9900 elite user manual software
Additional Information
Ge oec 9900 training manual

Additional Information

Specifications

X-ray System

Generator
• 60kHz high frequency
• 15kW power
• Up to 120kVp
• Up to 75mA for radiographic film exposure
• Continuous high level fluoro (HLF) up to 20mA
• Digital spot up to 75mA
• Digital cine pulse
- 15 or 30 pulses per second, 60 Hz
- 12 or 25 pulses per second, 50 Hz
- Up to 150mA
- 10ms pulse width
• Full power from standard wall outlet
• Patented battery buffered design


X-ray Tube
• Rotating anode X-ray tube

• 0.3 and 0.6 nominal focal spots
• Anode heat capacity: 300,000 HU
• Anode cooling rate: 85,000 HU/min.
• Housing heat capacity: 1,600,000 HU
• Housing cooling rate:
- 9” I.I. Super C- 22,500 HU/min
- 12” I.I. - 22,500 HU/min


Digital Image Rotation
• Digitally adjusts image display
• Automatic image update
- Image rotation
- Image reversal (side-to-side)
- Image invert (top-to-bottom)
• Image positioning without additional exposure


PreView Collimator
• On-screen collimator position indication
• PreView iris collimator
• PreView Tungsten rotatable double leaf collimator
• Adjusts collimators without X-ray exposure

Fluoro Mode
• kVp range: 40 -120
• mA range: 0.2 - 10 normal mode 1.0 - 20 HLF (high level fluoro)
• Auto and manual fluoro modes
• AutoTrak ABS varies mA, kVp, camera gain

Pulsed Fluoro Mode
• kVp range: 40 - 120
• mA range: 0.2 - 10
• Pulse rate: 1, 2, 4, 8
• Pulse width: 25 or 50ms
• AutoTrak ABS, mA, kVp, camera gain
• Reduces X-ray dose to patient and operator

High Level Pulsed Fluoro
• kVp range: 40 - 120
• mA range: 1 - 40
• Pulse rate: 1, 2, 4, 8
• Pulse width: 25 or 50ms
• AutoTrak ABS, mA, kVp, and camera gain

Digital Spot Mode
• kVp range: 40 - 120
• mA range: Up to 75
• Automatic exposure termination and automatic image save Radiographic Mode
• mA range: up to 75
• mAs range: up to 300
• Computer controlled exposure time
• Optional film cassette holder
- 10” x 12” (24cm x 30cm) for 9” I.I.
- 14” x 14” (35cm x 35cm) for 12” I.I.

Video Imaging System 9” Image Intensifier
• Tri-mode 9”/6”/4.5” (23cm/15cm/11cm) image intensifier
• Minimum central resolution (at monitor):
- 9” (23cm): 2.2 lp/mm
- 6” (15cm): 3.0 lp/mm
- 4.5” (11cm): 3.5 lp/mm
• DQE: 65% (typical) 12” Image Intensifier
• Tri-mode 12”/9”/6” (31cm/23cm/15cm) image intensifier
• Minimum central resolution (at monitor):
- 12” (31cm): 1.6 lp/mm
- 9” (23cm): 2.2 lp/mm
- 6” (15cm): 2.6 lp/mm
• DQE: 65% (typical)


Precision Imaging with Dynamic Range Management (DRM)
• Preset Imaging Profile: Orthopedic
• Enhances features of interest while attenuating background noise

AutoTrak Automatic Brightness Stabilization (ABS)
• Automatically seeks the subject anatomy anywhere within the imaging field and selects the optimum imaging technique
• Automatically adjusts to anatomical size and location
• Provides uniform image quality throughout entire image
• Simplifies operation

Image Quality Features
• Smart Window
- Dynamically senses the collimator position and automatically adjusts brightness and contrast to produce high image quality
• Smart Metal
- Allows user to adjust automatic brightness and contrast sensitivity levels to metal
- Provides optimum image quality even when metal is introduced to the field
• Tungsten Collimator
- Dense collimator limits X-ray exposure area
- Reduces scatter radiation
- Improves image detail Video Camera
• High resolution 1k x 1k CCD camera
• Full frame capture
• Motorized rotation
• On-screen orientation indicator (real-time feedback without fluoro)
• Left-right image reversal
• Top-bottom image invert Video Display
• Dual 19” (48cm) display anti-glare, LCD flat panel monitors mounted on an articulating arm
- 22” horizontal travel
- 7° up/10° down
- Monitors viewable from all four sides of workstation
- Horizontal and vertical viewing angle 170°

• 1200 CD/M2 maximum brightness
• Touch screen system control
• 1280 x 1024 high resolution monitors
• Integrated color monitor for display of VGA, DVI, DVI-D, S-VHS, and SDISD formats

Software Package Office Pain (OP) and Ambulatory Center (AC)
• 1k x 1k x 16 bit image processing
• Preset Imaging Profile: Orthopedic
• Noise filter with on-screen indicator
• Minimal difference spatiotemporal noise filter (MDST)
• Real-time dynamic range management (DRM)
• Automatic digital brightness and contrast control
• Manual digital brightness and contrast control
• Negate mode
• Save and auto-save feature
• Swap and auto-swap feature
• Patient information
- Examination list
- Customized patient information
• Customize functions
- Workstation set-up
- Mainframe set-up
- Patient information set-up
- Date/time set-up
- DICOM interface set-up
• Last image hold
• 63 image storage
• CD/DVD burner with DICOM viewer for displaying images on PC platforms
- 512 x 512 or 1k x 1k
• Integrated DICOM interface (store, print, worklist, and query/retrieve)
• HIPAA SecureView
- Password protection
- Blank screen function
- Delete all


Ambulatory Center Extended (ACE)
Includes all the AC features and:
• 1000 image storage
• Zoom and roam function
• Image annotation
• Measurement software


Office Pain Extended (OPE)
Includes all the OP features and:
• Real-time digital subtraction (DSA)
• Peak opacification
• 8fps Dynamic digital disk
- Recording/playback rate: 1, 2, 4, 8fps
- Recording time: 5 minutes @ 8fps
- Automatic image playback
- Frame-by-frame review

Additional Features
Hardcopy Options • Integrated film/paper printer
- No film developing required
- 8” x 10” (20.3cm x 25.5cm) laser quality film/paper
- Multi-format, 1, 2, 4 on 1
- Multi-copy capability
• Thermal printers
• Integrated DICOM interface (storage class/print class/query work-list)
User Interface
• Entire system is computer controlled
• Touchscreen control simplifies operation
• Automated system operation requires minimum operator interface

• Multi-functional controls
- Footswitch
- IR remote (optional)
- Hand-held control
• Simplified keyboard with integrated touchpad
• Multi-purpose image directory
- Retrieve and review images
- Compose hardcopy films
- Copy images
• X-ray dose summary

Uninterruptible Power
• Orderly shutdown Electrical
• Input power (60 Hz or 50 Hz) 9900 Elite
- 100V @ 20 A
- 120V @ 12 A
Regulatory Compliance
• U.S. 21 CFR Subchapter J
• NFPA 99
• UL 60601-1 (CSA/NRTL)
• IEC60601-1 (plus relevant Collateral and Particular Standards)

File AttachmentsNo